Type of Document Master's Thesis Author Hundley, Robert Wynne URN etd-03302010-020116 Title Improvements to the kidney dialysis machine Degree Master of Science Department Mechanical Engineering Advisory Committee
Advisor Name Title Hurst, Charles J. Committee Chair Moses, Hal L. Committee Member Reinholtz, Charles F. Committee Member Keywords
- Hemodialysis
Date of Defense 1992-02-05 Availability restricted Abstract Five changes were suggested to improve the kidney dialysis machine, including a conductivity control feedback system, a positive pressure deaeration system, a balancing system for dialysate flow control, a diaphragm-type blood pump, and an ultrafiltration control system using blood pressure measurement. A kidney dialysis machine was constructed to test the changes for possible use in future dialysis machine designs. Tests were made on each of the five system improvements. All five ideas had some merit for a commercial dialysis machine. A description of the investigation and the various factors used in the development and testing of the five systems is included.
Files
Filename Size Approximate Download Time (Hours:Minutes:Seconds)
28.8 Modem 56K Modem ISDN (64 Kb) ISDN (128 Kb) Higher-speed Access LD5655.V855_1992.H862.pdf 2.78 Mb 00:12:52 00:06:37 00:05:47 00:02:53 00:00:14 next to an author's name indicates that all files or directories associated with their ETD are accessible from the Virginia Tech campus network only.
If you have questions or technical problems, please Contact DLA.